Energy-based Devices

 

ablation zone examples

The following are examples of potential ablation results using single and multiple Evident™ MWA Percutaneous Antennas to coagulate tissue prior to resection.

These examples are from ex vivo animal tissue (liver) models. The size of ablation zones may vary in clinical settings.

Ablation Zones of Planar Coagulation

Single Evident™ MWA Percutaneous Antenna Configuration

Single Percutaneous MWA Antenna Configuration for Planar Coagulation

45 W timer icon 1:00 timer icon 2:00
A 0.8 cm 1.2 cm
B 2.8 cm 3.4 cm

Linear Evident™ MWA Percutaneous Antennas Configuration

Linear MWA Percutaneous Antenna Configuration for Planar Coagulation

45 W timer icon 2:00 2 antennas 3 antennas 4 antennas
S 1.0 cm 1.0 cm 1.0 cm
A 2.5 cm 3.4 cm 4.3 cm
B 3.6 cm 3.6 cm 3.6 cm
C 1.9 cm 1.9 cm 1.9 cm

 

Cluster Evident™ MWA Percutaneous Antennas Configuration

Cluster MWA Percutaneous Antenna Configuration for Planar Coagulation

45 W timer icon 0:30
S 0.5 cm
A 1.1 cm
B 3.2 cm

Evident™ MWA System Tissue Verification Testing Protocol
Ablation zone is dependent upon tissue vascularization, type, temperature and impedance.
Ablation zone testing was performed in healthy ex vivo bovine 20°C liver. Ablation zones may vary in clinical settings.
Illustrations of ablation zones are not to scale.