Energy-based Devices

 

ablation zone examples

The following are examples of potential ablation results using single and multiple MWA surgical antennas.

These examples are from ex vivo animal tissue (liver) models. The size of ablation zones may vary in clinical settings.

Ablation Zones

Single Evident™ MWA Surgical Antenna

Single Surgical MWA Antenna Configuration

timer icon 45 W
10:00
A 2.7 cm
B 4.1 cm
M 0.6 cm
Volume 16 cm3

 

Two Evident™ MWA Surgical Antennas

Two MWA Surgical Antenna Configuration

 

  45 W timer icon 10:00
S 1.5 cm 2.0 cm
A 4.2 cm   4.5 cm
B 4.5 cm   4.7 cm
C 3.4 cm   3.3 cm
M 0.7 cm   0.7 cm
Volume 34 cm3   37 cm3

 

Three Evident™ MWA Surgical Antennas

Three MWA Surgical Antenna Configuration

 

  45 W timer icon 10:00
S 1.5 cm 2.0 cm
A 4.1 cm   4.6 cm
B 4.6 cm   5.1 cm
M 0.8 cm   0.8 cm
Volume 40 cm3   56 cm3

Evident™ MWA System Tissue Verification Testing Protocol
Ablation zone is dependent upon tissue vascularization, type, temperature and impedance.
Ablation zone testing was performed in healthy ex vivo bovine 20°C liver. Ablation zones may vary in clinical settings.
Illustrations of ablation zones are not to scale.