Principles of Electrosurgery

Properties of Electricity

Principles of Electrosurgery

Bipolar Electrosurgery

Monopolar Electrosurgery

Tissue Effects Change as You Modify the Waveform

Electrosurgical Tissue Effects

Grounded Electrosurgical Systems

Isolated Electrosurgical Systems

Patient Return Electrodes

Function of the Patient Return Electrode

Ideal Return Electrode Contact with Current

Assess Pad Site Location

Patient Return Electrode Monitoring Technology

Adaptive Technologies

Electrosurgery Safety Concerns During MIS

Capacitive Coupling During Endoscopy

Recommendations to Avoid Electrosurgical Patient Complications in MIS

Coated Electrodes

Argon-Enhanced Electrosurgery

Surgical Smoke

OR Safety Precautions

Bibliography

Patient Return Electrodes

Function of the Patient Return Electrode

The function of the patient return electrode is to remove current from the patient safely. A return electrode burn occurs when the heat produced, over time, is not safely dissipated by the size or conductivity of the patient return electrode.

BURN = (Current x Time) / Area